innovatec.pune@yahoo.com |
+91-9011002697 |
Specifications
Model | VEGADIF 65 |
Process-pressure | -1 - 420 bar |
Threaded-connection | ¼ - 18 NPT |
Flange-connection | ≥ DN32, ≥ 1⅜ |
Ambient-temperature | -40 - 80 °C |
Measuring-range Pressure | -40 - 40 bar |
Smallest adjustable span | 0.25 mbar/25 Pa (0.036 psig) |
Deviation in characteristics | 0.15 %; 0.075 % |
Process fitting - basic version | ¼-18 NPT according to IEC 61518 |
Process fitting with chemical seals | Flanges from DN 32 or 2 |
Process temperature - basic version | -40 - +120 °C (-40 to +248 °F) |
Process temperature with chemical seals | -40 - +400 °C (-40 - +752 °F) |
Process pressure | -1 - +420 bar/-100 - +42000 kPa (-14.5 - +6092 psig) |
Operating voltage | 12 - 36 V DC |
Application area : The differential pressure transmitter VEGADIF 65 can be used universally for the measurement of liquids, gases and vapours. Typical applications are level measurements in pressurized vessels as well as flow measurements in combination with DP flow elements. Further applications are pressure monitorings on filters as well as density and interface measurements.
Materials, wetted-parts
Seal-material
Housing-material
Protection-rating
Output
Your benefit
Function : The sensor diaphragms are deflected by the acting pressures. The filling oil transmits the pressure to a semiconductor measuring bridge. The output signal, which is dependent on the differential pressure, is measured and further processed.
Materials : The process fitting is made of steel C22.8 (1.0460), Alloy or stainless steel 316L. The process diaphragm is available in 316L and high resistance materials Alloy C276 (2.4819), Alloy 400 (2.4360), Tantalum as well as Rhodium/Gold on 316L.
Housing versions : The housings are available as single chamber or double chamber version in plastic, stainless steel or aluminium. They are available in protection ratings up to IP 68 (25 bar) with external electronics.
Electronics versions : The instruments are available in different electronics versions. Apart from the two-wire electronics with 4 - 20 mA or 4 - 20 mA/HART, two purely digital versions with Profibus PA and Foundation Fieldbus are available.